EasyManuals Logo
Home>Tektronix>Test Equipment>MSO56

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
Commands listed in alphabetical order
Arguments
<NR1> =0stopsa
cquisitions; any other value starts acquisitions.
OFF stops acquisitions.
ON starts acquisitions.
RUN starts acquisitions.
STOP stops acquisitions.
Examples
ACQUIRE:STATE RUN starts the acquisition of waveform data and resets the count
of the number of acquisitions.
ACQUIRE:STATE? might return :ACQUIRE:STATE 0, indicating that the
acquisition is stopped.
ACQuire:STOPAfter
This command sets or queries whether the instrument continually acquires
acquisitions or acquires a single sequence. Pressing SINGLE on the front
panel button is equivalent to sending these c ommands:
ACQUIRE:STOPAFTER
SEQUENCE
and ACQUIRE:STATE 1.
Group
Acquisition
Syntax
ACQuire:STOPAfter {RUNSTop|SEQuence }
ACQuire:STOPAfter?
Related Commands
ACQuire:STATE
Arguments
RUNSTop species that the instrument will continually acquire data, if
ACQuire:STATE is turned on.
SEQuence species that the next acquisition will be a single-sequence acquisition.
Examples
ACQUIRE:STOPAFTER RUNSTOP sets the instrument to continually acquire data.
ACQUIRE:STOPAFTER? might return :ACQUIRE:STOPAFTER SEQUENCE,
indicating that the next acquisition the instrument makes will be of the
single-sequence type.
AFG:AMPLitude
Sets (or queries) the AFG a mplitude in volts, peak to peak.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-99