Commands listed in alphabetical order
EOW specifies tr
iggering on the 32nd bit of a word.
ERRor specifies triggering on a specified error condition.
NOTE. The typ
e of error triggered on is specified by
TRIGger:{A |B}:BUS:B<x>:ARINC429A:ERRTY Pe
.
Examples
TRIGGER:A:BUS:B1:ARINC429 A:CONDITION DATA specifies triggering on
packets that contain matching data field(s).
TRIGGER:A:BUS:B1:ARINC429 A:CONDITION? might return SOW, indicating
that the bus is triggering on the first bit of each word in the pacekt.
TRIGger:{A|B}:BUS:B<x>:ARINC429A:DATa:HIVALue
This command sets or queries the high value when trigger on an ARINC429 data
field. The bus number is specified by x. The trigger condition must be set to DATa
or LABELANDDATA, and the data qualifier must be INrange or OUTrange.
Conditions
Requires the SR-AERO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:AR INC429A:DATa:HIVALue <QString>
TRIGger:{A|B}:BUS:B<x>:AR INC429A:DATa:HIVALue?
Arguments
<QString> is the label value.
NOTE. The size of the QString is dependent on the data field format selected
using BUS:Bx:ARINC429A:DATAFORmat. Also, the stored QString is reset to its
default value whenever the data field format is changed.
Examples
TRIGGER:A:BUS:B1:ARINC429 A:DATA:HIVALUE
"XXXXXXXXXXXXXXX1000"
sets the value to XXXXXXXXXXXXXXX1000.
TRIGGER:A:BUS:B1:ARINC429 A:DATA:HIVALUE? might return
"XXXXXXXXXXXX XXXXXXX", indicating that the value is
XXXXXXXXXXXXXXXXXXX.
2-760 MSO54, MSO56, MSO58, MSO58LP Programmer