Commands listed in alphabetical order
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:SPI :CONDition {SS|STARToffra me|DATa}
TRIGger:{A|B}:BUS:B<x>:SPI :CONDition?
Related Commands
BUS:B<x>:SPI:IDLETime,
BUS:B<x>:SPI:FRAMING
Arguments
SS specifies the Slave Selection condition.
STARTofframe is applicable when BUS:B<x>:SPI:FRA
MING
is set to
IDLEtime. When the trigger condition is set to STARTofframe,theinstrument
triggers on the first SPI clock after an idle time when t here are no clocks.
DATa sets the trigger condition to Master-In Slave-Out and Master-Out Slave-In.
TRIGger:{A|B}:BUS:B<x>:SPI:DATa:SIZe
This command specifies the length of the data string to be used for a SPI trigger if
the trigger condition is DATa. The bus number is specified by <x>.
Conditions
Requires the SR-COMP Serial Triggering and Analysis Application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:SPI :DATa:SIZe <NR1>
TRIGger:{A|B}:BUS:B<x>:SPI :DATa:SIZe?
Arguments
<NR1> is the length of the data string in bytes.
Examples
TRIGger:A:BUS:B1:SPI:DATa: SIZe 1 sets the data size to 1 byte.
TRIGger:A:BUS:B1:SPI:DATa: SIZe? might return
:TRIGGER:A:BUS:B1:SPI:DATA :SIZE 1 indicating the data size is
1byte.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-833