Commands listed in alphabetical order
Conditions
Requires optio
n SR-EMBD.
Group
Bus
Syntax
BUS:B<x>:SPI:DATa:SOUrce
{CH<x>|CH<x>_D<x>|MATH<x>| REF<x>|REF<x>_D<x>}
BUS:B<x>:SPI:DATa:SOUrce?
Arguments
CH<x> designates an analog channel as the data source for the specied SPI bus.
CH<x>_D<x> designates an digital channel as the bus SPI clock source.
MATH<x> designates a math waveform as the data source.
REF<x> designates a reference waveform as the data source.
REF<x> _D<x> species a digital reference waveform as the data source
waveform for the specied SPI bus.
Examples
BUS:B1:SPI:DATA:SOURCE CH2 sets the SPI data source for the bus B1 to CH2.
BUS:B1:SPI:DATA:SOURCE? might return :BUS:B1:SPI:DATA: SOURCE CH3,
indicating that the SPI data source for the bus B1 is set to CH3.
BUS:B<x>:SPI:DATa:THReshold
This command sets or queries the SPI Data (SDA) source threshold for the
specied bus. The bus is specied by x.
Conditions
Requires option SR-EMBD.
Group
Bus
Syntax
BUS:B<x>:SPI:DATa:THReshol d <NR3>
BUS:B<x>:SPI:DATa:THReshol d?
Arguments
<NR3> is the SPI Data (SDA) source threshold for the specied bus.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-173

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals