Commands listed in alphabetical order
SEARCH:SEARCH
1:TRIGger:A:LOGIc:LEVel:CH2?
might return
:SEARCH:SEARCH1:TRIGGER:A :LOGIC:LEVEL:CH2 0.0E+0 indicating the
level is set to 0.0 V.
SEARCH:SEAR
CH<x>:TRIGger:A:LOGIc:LEVel:MATH<x>
This command sets the voltage level to use for logic trigger search. The search
number is specied by <x>.
Group
Search and Mark
Syntax
SEARCH:SEAR
CH<x>:TRIGger:A:LOGIc:LEVel :MATH<x> <NR3>
SEARCH:SEARCH<x>:TRIGger: A:LOGIc:LEVel:MATH<x>?
Arguments
<NR3> is the voltage level to use for logic trigger search.
Examples
SEARCH:SEARCH1:TRIGger:A: LOGIc:LEVel:MATH1 50.0e-3 sets the level
to 50.0 mV.
SEARCH:SEARCH1:TRIGger:A: LOGIc:LEVel:MATH1? might return
:SEARCH:SEARCH1:TRIGGER:A :LOGIC:LEVEL:MATH1 0.0E+0 indicating
the level is set to 0.0 V.
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:LEVel:REF<x>
This command sets the voltage level to use for logic trigger search. The search
number is specied by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:LOGIc:LEVel:REF<x> <NR3 >
SEARCH:SEARCH<x>:TRIGger: A:LOGIc:LEVel:REF<x>?
Arguments
<NR3> is the voltage level to use for logic trigger search.
Examples
SEARCH:SEAR
CH1:TRIGger:A:LOGIc:LEVel:R EF1 50.0e-3
sets the level
to 50.0 mV.
2-718 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals