Commands listed in alphabetical order
Arguments
1 or ON turns on a
utorange for the Digital Voltmeter.
0 or OFF turns autorange off.
Examples
DVM:AUTOR ON turns on autorange for the Digital Voltmeter.
DVM:MEASUrement:FREQuency? (Query Only)
This command returns the current frequency value for the DVM.
Conditions
Requires DVM option.
Group
DVM
Syntax
DVM:MEASUrement:FREQuency?
Related Commands
DVM:TRIGger:FREQuency:COUNTer
Examples
DVM:MEASU:FREQ? might return 100.0000E+3, which represents 100 kHz as
the current frequency value for the D VM.
DVM:MEASUrement:HIStory:AVErage? (Query Only)
Returns the average DVM readout value over the history period. The history
period is a constant period of 5 seconds.
Conditions
Requires DVM option.
Group
DVM
Syntax
DVM:MEASUrement:HIStory:AV Erage?
Examples
DVM:MEASU:HIS:AVE? might return :DVM:MEASUREMENT:HISTORY:AVERAGE
429.3000E-3
which represents the average DVM readout value over the history
period in volts.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-317