Commands listed in alphabetical order
MEASUrement:R
EF1:REFLevels:ABSolute:TYPE?
might return
:MEASUREMENT:REF1:REFLEVEL S:ABSOLUTE:TYPE SAME indicating that the
absolute levels are set the same.
MEASUrement:REF<x>:REFLevels:BASETop
This command sets or queries the method used to calculate the TOP and BASE
,
used to calculate reference levels for the measurement.
Group
Measurement
Syntax
MEASUrement:REF<x>:REFLeve ls:BASETop
Arguments
AUTO automatically chooses a reference level method.
MINMax specifies that reference levels a re relative to the measurement MIN and
MAX.
MEANhistogram specifies that reference levels are relative to the histogram mean
BASE and TOP.
MODEhistogram specifies that reference levels are relative to the histogram mode
BASE and TOP.
EYEhistogram specifies that reverence levels are relative to the eye histogram
BASE and TOP.
Examples
MEASUrement:REF1:REFLevels :BASETop MINMAX specifies that reference
levels are relative to the measurement MIN and MAX.
MEASUrement:REF1:REFLevels :BASETop? might return
:MEASUREMENT:REF1:REFLEVEL S:BASETOP AUTO indicating the instrument
automatically chooses a reference level method.
MEASUrement:REF<x>:REFLevels:METHod
Thi
s command sets or queries the method used to calculate reference levels for
the measurement.
Gr
oup
Me
asurement
Syntax
MEASUrement:REF<x>:REFLeve ls:METHod {PERCent|ABSolu te}
MEASUrement:REF<x>:REFLeve ls:METHod?
MSO54, MSO56, MSO58, MSO58LP Programmer 2-511