Commands listed in alphabetical order
Examples
MEASUREMENT:C
LOCKRECOVERY:DATAPATH
"C:/E:" sets the data path to C:/E:.
MEASUREMENT:CLOCKRECOVERY: DATAPATH? might return
:MEASUREMENT
:CLOCKRECOVERY:DATAPATH "C:/E"
.
MEASUrement:CLOCKRecovery:DATARate
This command sets or queries the global nominal data bit rate used when
nominal data rate clock recovery is used for the measurement. This will affect
measurements whose :MEASUrement:MEAS<x>:CLOCKRecovery:GLOBal
flag is set to 1.
Group
Measurement
Syntax
MEASUrement:CLOCKRecovery: DATARate <NR3>
MEASUrement:CLOCKRecovery: DATARate?
Arguments
<NR3> is the global value for the Nominal data rate.
Examples
MEASUREMENT:CLOCKRECOVERY: DATARATE 2.0e+9 setsthedatarateforclock
recovery to 2.0 GHz.
MEASUREMENT:CLOCKRECOVERY: DATARATE? might return
:MEASUREMENT:CLOCKRECOVERY :DATARATE 2.5000E+9 indicating
the data rate is 2.5 GHz.
MEASUrement:CLOCKRecovery:EXPLICITCLOCKMODe
This command sets or queries the global explicit clock mode used when explicit
clo
ck recovery is used for the measurement. This w ill affect measurements whose
:MEASUrement:MEAS<x>:CLOCKRecovery:GLOBal flag is set to 1.
Group
Measurement
Syntax
MEASUrement:CLOCKRecovery: EXPLICITCLOCKMODe {EDGE|P LL}
MEASUrement:CLOCKRecovery: EXPLICITCLOCKMODe?
Arguments
EDGE sets the clock mode to clock edge.
PLL sets the clock mode to phase locked loop.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-395