EasyManuals Logo
Home>Tektronix>Test Equipment>MSO56

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #368 background imageLoading...
Page #368 background image
Commands listed in alphabetical order
HORizontal:M
ODE:SAMPLERate
This command sets or queries the sample rate.
Group
Horizontal
Syntax
HORizontal:MODE:SAMPLERat e <NR1>
HORizontal:MODE:SAMPLERat e?
Arguments
<NR1> is the sample rate in samples per se cond.
Examples
HORIZONTAL:MODE:SAMPLERAT E 1e6 sets the sample rate to 1 million samples
per second.
HORIZONTAL:MODE:SAMPLERAT E? might return
:HORIZONTAL:MODE:SAMPLERA TE 5.0000E+6, indicating that the sample rate
is set to 5 million samples per second.
HORizontal:MODE:SCAle
This command sets or queries the horizontal scale.
Group
Horizontal
Syntax
HORizontal:MODE:SCAle <NR1>
HORizontal:MODE:SCAle?
Arguments
<NR1> is the horizontal scale in seconds per division.
Examples
HORIZONTAL:MODE:SCALE 2e-9 sets the horizontal scale to 2 ns per division.
HORIZONTAL:MODE:SCALE? might return :HORIZONTAL:MOD E:SCALE
20.0000E-6
, indicating that the horizontal scale is set to 10 μs per division.
HORizontal:POSition
This command s ets or queries the horizontal position as a percent of screen
width. When Horizontal Delay Mode is turned off, this command is equivalent
to adjusting the HORIZONTAL POSITION knob o n the front panel. When
Horizontal Delay Mode is turned on, the horizontal position is forced to 5 0%.
2-352 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals