Commands listed in alphabetical order
Group
Horizontal
Syntax
HORizontal:FASTframe:REF: INCLUde {<NR1>|OFF|ON}
HORizontal:FASTframe:REF: INCLUde?
Arguments
ON displays the delta information.
OFF does not display the delta information.
<NR1> a 0 indicates the delta information is off; any other value displays the
delta information.
HORizontal:FASTframe:SELECTED
This command sets or returns the selected frame number for acquired frames.
Refs have their own selected frames.
Group
Horizontal
Syntax
HORizontal:FASTframe:SELE CTED <NR1>
HORizontal:FASTframe:SELE CTED?
Arguments
<NR1> is the selected frame number for acquired frames.
Examples
HORIZONTAL:FASTFRAME:SELE CTED? might return
:HORIZONTAL:FASTFRAME:SEL ECTED 1, indicating the selected frame number
is 1.
HORizontal:FASTframe:STATE
This command sets or returns the state of FastFrame. Acq
uisition modes Envelope
and Average are n ot compatible with FastFrame. If FastFrame is on, an attempted
set to those acquisition m odes will fail and revert to Sample mode. If FastFrame
is turned on while in one of those acquisition modes, the acquisition mode is
changed to Sample.
Group
Horizontal
Syntax
HORizontal:FASTframe:STAT E {<NR1>|OFF|ON}
HORizontal:FASTframe:STAT E?
2-344 MSO54, MSO56, MSO58, MSO58LP Programmer