Commands listed in alphabetical order
MATH<x> =Amath
waveform specifier; <x> is ≥1.
REF<x> = A reference waveform specifier; <x> is ≥1.
Examples
POWer:POWer1:CYCAmp:INPUT SOurce CH1 sets the input source for cycle
amplitude measurement as CH1 for the power measurement badge Power 1.
POWer:POWer<x>:CYCLEBase:INPUTSOurce
This command sets or queries the input source for cycle base measurement in the
specified power measurement badge. The power measurement badge is specified
by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:CYCLEBase: INPUTSOurce {CH<x> | MATH<x> | REF<x>}
POWer:POWer<x>:CYCLEBase: INPUTSOurce?
Arguments
CH<x> = A channel specifier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x> = A math waveform specifier; <x> is ≥1.
REF<x> = A reference waveform specifier; <x> is ≥1.
Examples
POWer:POWer1:CYCLEBase:IN PUTSOurce CH2 sets the input source for cycle
base measurement as CH2 for the power measurement badge Power 1.
POWer:POWer<x>:CYCLEMAX:INPUTSOurce
This command sets or queries the input source for cycle maximum measurement
in the specified power measurement badge. The power measurement badge
is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
2-530 MSO54, MSO56, MSO58, MSO58LP Programmer