Appendix D: Waveform Transfer (WFMOutpre and CURVe Query) Examples
WFMOutpre? Query results
Explanation
NR_PT 25
This value indicates the number of data points in the waveform record to be transferred
using the CURVE? query. (If you would like to determine only this value, use the
WFMOutpre:NR
_Pt? query.) Note: this value is typically equal to the full record length,
but you also have the option to transfer only a portion of the record length by using the
DATa:STARt and DATa:STOP commands.
PT_FMT Y
This value indicates the format of the data points in the waveform record. In this c ase, the
value repre
sents YT format. This is query only. (If you would like to determine only this
value, use the WFMOutpre:PT_Fmt? query.)
PT_ORDER L
INEAR
This value i
salways
LINear.
XUNIT "s"
This value
indicates the units of the x-axis of the waveform record. This is determined by
the horizontal settings for the waveform source. Typically, this value is “s”, representing
seconds. When using the math waveform as a source, the value can be “s” or “Hz”. This
is query o
nly. (If you would like to determine only this value, use the WFMOutpre:XUNit?
query.)
XINCR 100
.0000E-9
This value indicates the time, in seconds, or frequency, in hertz, between data points i n the
waveform record. This is query only. (If you would like to determine only this value, use
the WFMO
utpre:XINcr? query.)
XZER0 -500.0000E-6
This va
lue indicates the time, in seconds, or frequency, in hertz, of the rst data point in the
waveform record. This time or frequency is relative to the time of the trigger, which is always
0. So, this XZEro time or frequency can be negative. This is query only. (If you would like to
determ
ine only this value, use the WFMOutpre:XZEro? query.)
PT_OFF 0
This i
s a query provided only for compatibility with other Tektronix oscilloscopes. The
returned value is always 0. (If you would like to determine only this value, use the
WFMOutpre:PT_Off? query.)
YUNIT "State"
This value indicates the vertical units of data points in the waveform record. This can be
any o
f several string values, depending upon the vertical units of the source waveform
in this case, State. This is query only. (If you would like to determine only this value, use
the WFMOutpre:YUNit? query.)
YMULT 1.0000
This value indicates the multiplying factor to convert the data point values from digitizing
lev
els to the units specied by the
WFMOUTpre:YUNit command. This is query only. (If
you would like to determine only this value, use the WFMOutpre:YMUlt? query.)
YO
FF 0.0E+0
This value indicates the vertical position of the source waveform in digitizing levels. There
are 25 d igitizing levels per vertical division. This is query only. (If you would like to determine
on
ly this value, use the WFMOutpre:YOFf? query.)
YZERO 0.0E+0
T
his value indicates the vertical offset of the source waveform in units specied by the
WFMOUTpre:YUNit command. This is query only. (If you would like to determine only
this value, use the WFMOutpre:YZEro? query.)
Example 3: Digital with 4 Bytes Per Point and Zoom Off
Goal: Transfer 25 points of Digital data from the oscilloscope to a PC using 4
bytes per point and Zoom off.
MSO54, MSO56, MSO58, MSO58LP Programmer D-5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals