Commands listed in alphabetical order
Group
Display Contro
l
Syntax
DISplay:GLObal:CH<x>:STATE {<NR1>|OFF| ON}
Arguments
<NR1> = 0 disables the display of the specified channel; any other value enables
display of the channel.
ON enables display of the specified channel.
OFF disables display of the specified channel.
Examples
DISplay:GLObal:CH1:STATE OFF turns off the display of channel 1.
DISplay:GLObal:CH1:STATE? might return :DISPL AY:GLOBAL:CH1:STATE
0
indicating that channel 1 is not displayed.
DISplay:GLObal:MATH<x>:STATE
This command sets or queries the global state (display mode On or Off) of the
specified math. Setting this value true (On or NR1 ≠ 0 ) turns on the source in the
waveform view. Setting this value false (Off or NR1 = 0 ) turns off the source in
the waveform view. This command only works if the specified math waveform
is added already.
Group
Display Control
Syntax
DISplay:GLObal:MATH<x>:STA TE {<NR1>|OFF|ON}
Arguments
<NR1> = 0 disables the display of the specified math; any other value enables
display of the math.
ON enables dis play of the specified math.
OFF disables display of the specified math.
Examples
DISplay:GLObal:MATH1:STATE 1 displays math 1.
DISplay:GLObal:MATH1:STATE ? might return
:DISPLAY:GLOBAL:MATH1:STAT E 1 indicating that math1 is displayed.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-229