Commands listed in alphabetical order
Group
Trigger
Syntax
TRIGger:{A|B}:WINdow:WIDth <NR3>
Arguments
<NR3> is the minimum width for a window violation.
Examples
TRIGger:A:WINdow:WIDTH 4.5e-9 sets the minimum width for a window
violation to 4.5 ns.
TRIGger:A:WINdow:WIDTH? might return :TRIGGER:A:WINDOW:WIDTH
4.0E-9
indicating that 4.0 ns is the minimum width for a window violation.
TRIGger:A:HOLDoff:BY
This command sets or q
ueries the type of holdoff for the A trigger. Holdoff types
are expressed as either user-specied time (TIMe) or by an internally calculated
random time value (RANDom). This command is equivalent to selecting Mode &
Holdoff from the Trig menu and then setting the Holdoff type.
Group
Trigger
Syntax
TRIGger:A:HOLDoff:BY {TIMe|RANDom}
TRIGger:A:HOLDoff:BY?
Related Commands
TRIGger:A:HOLDoff:TIMe
Arguments
TIMe enables you to set the holdoff time via the TRIGger:A:HOLDoff:TIMe
command.
RANDom species a random time value.
Examples
TRIGGER:A:HOLDOFF:BY TIME sets the holdoff to the "by time" setting. This
enables you to set the holdoff time.
TRIGGER:A:HOLDOFF:BY? might return :TRI GGER:A:HOLDOFF: BY TIME,
indicating that you will set the holdoff time.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-875