EasyManuals Logo
Home>Tektronix>Test Equipment>MSO56

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #907 background imageLoading...
Page #907 background image
Commands listed in alphabetical order
type. Some wave
forms are normalized vector data where the data points are
8-byte doubles in oating p oint format whereas other formats are 1-byte or 2-byte
integers.
Group
Waveform Transfer
Syntax
WFMOutpre:ASC_Fmt?
Related Commands
DATa:SOUrce, WFMOutpre:BN_Fmt, WFMOutpre:ENCdg
Returns
FP repres ents oating point ASCII data. The waveforms are normalized vector
data where the data points are 8-byte doubles in oating point format.
INTEGER represents signed integer ASCII data. The waveform data are 1-byte
or 2-byte integers.
Examples
WFMOutpre:ASC_Fmt? might return :WFMOUTPRE:ASC_FMT INTEGER
indicating the ASCII format is integer.
WFMOutpre:BIT_Nr
This command sets and queries the number of bits per waveform point that
outgoing waveforms contain, as specied by the DATa:SOUrce command. Note
that values will be constrained according to the underlying waveform data. This
specication is only meaningful when WFMOutpre:ENCdg is set to BIN and
WFMOutpre:BN_Fmt is set to either RI or RP.
Group
Waveform Transfer
Syntax
WFMOutpre:BIT_Nr <NR1>
WFMOutpre:BIT_Nr?
Related Commands
DATa:SOUrce, WFMOutpre:BN_Fmt, WFMOutpre:ENCdg
Arguments
<NR1> number of bits per data point can be 8 or 16.
Examples
WFMOUTPRE:BIT_NR 16 sets the number of bits per waveform point to 16 for
incoming RI and RP binary format data.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-891