Commands listed in alphabetical order
Syntax
WFMOutpre:YMU
lt?
Related Commands
DATa:SOUrce
Examples
WFMOUTPRE:YMULT? might return :WFMOUTPRE:YMULT 4.0000E-3, indicating
that the vertical scale for the corresponding waveform is 100 mV/div.
WFMOutpre:YOFf? (Query Only)
This query-only command returns the vertical offset of the source specied by
DATa:SOUrce. For this instrument family, the valuereturnedisalways0.0asthe
offset is combined with the :YZEro value.
An error is reported if the DATa:SOUrce waveform does not exist.
Group
Waveform Transfer
Syntax
WFMOutpre:YOFf?
Related Commands
DATa:SOUrce, WFMOutpre:BYT_Nr
Examples
WFMOUTPRE:YOFF? might return :WFMOUTPRE:YOFF -50.0000E+0, indicating
that the position indicator for the waveform was 50 digitizing levels (2 divisions)
below center screen.
WFMOutpre:YUNit? (Query Only)
This query-only command returns the vertical units for the waveform specied by
the DATa:SOUrce command.
An error is reported if the DATa:SOUrce waveform does not exist.
Group
Waveform Transfer
Syntax
WFMOutpre:YUNit?
Related Commands
DATa:SOUrce
MSO54, MSO56, MSO58, MSO58LP Programmer 2-899