EasyManuals Logo
Home>Tektronix>Test Equipment>MSO56

Tektronix MSO56 User Manual

Tektronix MSO56
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #278 background imageLoading...
Page #278 background image
Commands listed in alphabetical order
Examples
DISplay:PLOTV
iew1:ZOOM:XAXIS:TO 2.5
sets the end of the zoom x-axis
to 2.5 V.
DISplay:PLOT
View3:ZOOM:XAXIS:TO?
might return
:DISPLAY:PLOTVIEW3:ZOOM:X AXIS:TO 2.2144883507508
indicating the end of the zoom x-axis is 2.21 V.
DISplay:PL
OTView<x>:ZOOM:YAXIS:FROM
This command sets or queries the bottom value of the zoom y-axis in the specied
plot view.
Group
Zoom
Syntax
DISplay:PLOTView<x>:ZOOM: YAXIS:FROM <NR3>
DISplay:PLOTView<x>:ZOOM: YAXIS:FROM?
Arguments
<NR3> is the bottom value of the zoom y-axis.
Examples
DISplay:PLOTView1:ZOOM:YA XIS:FROM -1.0E+20 sets the bottom value
of the zoom y-axis to -1E+20.
DISplay:PLOTView1:ZOOM:YA XIS:FROM? might return
:DISPLAY:PLOTVIEW1:ZOOM:Y AXIS:FROM -1.0E+21 indicating
the bottom value of the z oom y-axis is -1E+21.
DISplay:PLOTView<x>:ZOOM:YAXIS:TO
This command sets or queries the top value of the zoom y-axis in the specied
plot view.
Group
Zoom
Syntax
DISplay:PLOTView<x>:ZOOM: YAXIS:TO <NR3>
DISplay:PLOTView<x>:ZOOM: YAXIS:TO?
Arguments
<NR3> is the top value of the zoom y-axis.
Examples
DISplay:PLOTView1:ZOOM:YA XIS:TO 100 sets the top value of the zoom
y-axis to 100.
2-262 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals