Commands listed in alphabetical order
EQual argument
sets the oscilloscope to search when the pattern
is true for a time period equal to the time period specified in
SEARCH:SEARCH<x>:TRIGger: A:RUNT:WIDth within a ±5% tolerance.
NOTEQual argument sets t he oscilloscope to search w hen the pattern is true for
atime period greater than or less than (but not equal) the time period specified in
SEARCH:SEARCH<x>:TRIGger: A:RUNT:WIDth within a ±5% tolerance.
OCCURS argument specifies a search event if a runt of any detectable width occurs.
Examples
SEARCH:SEARCH1:TRIGGER:A: RUNT:WHEN MORETHAN sets the oscilloscope to
trigger when a runt pulse is detected with width wider than the time set by the
SEARCH:SEARCH<x>:TRIGger:A:RUNT:WIDth command.
SEARCH:SEARCH1:TRIGGER:A: RUNT:WHEN? might return
:SEARCH:SEARCH1:TRIGGER:A :RUNT:WHEN OCCURS, indicating
thatamarkisplacedifarunt trigger event occurs.
SEARCH:SEARCH<x>:TRIGger:A:RUNT:WIDth
This command sets or queries the width setting for a runt trigger search to
determine where to place a mark. The search number is specified by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:RUNT:WIDth <NR3>
SEARCH:SEARCH<x>:TRIGger: A:RUNT:WIDth?
Arguments
<NR3> specifi es the minimum width in seconds.
Examples
SEARCH:SEARCH1:TRIGGER:A: RUNT:WIDTH 400E–12 sets the runt trigger
minimum width for search 1 to 0.4 nanoseconds.
SEARCH:SEARCH1:TRIGGER:A: RUNT:WIDTH? might return
:SEARCH:SEARCH1:TRIGGER:A :RUNT:WIDTH 500.00O0E-12, indicating that
the runt trigger minimum width for search 1 is set to 0.5 nanoseconds.
SEARCH:SEARCH<x>:TRIGger:A:SETHold:CLOCk:EDGE
This command sets or queries the clock slope setting for a setup/hold trigger
search to determine where to place a mark. The search number is specified by <x>.
Group
Search and Mark
2-730 MSO54, MSO56, MSO58, MSO58LP Programmer