Commands listed in alphabetical order
Group
Vertical
Syntax
CH<x>:PROBECOntrol {AUTO|MANual}
CH<x>:PROBECOntrol?
Arguments
AUTO setsthevalues.Theproberangeis automatically calculated.
MANual allows you to select various valid values for the probe connected to a
particular channel.
Examples
CH2:PROBECONTROL AUTO sets the values and the probe range is automatically
calculated.
CH2:PROBECONTROL? might return :CH2:PROBECONTROL MANUAL indicating
that you c an select various valid values for the probe connected to channel 2.
CH<x>:PROBEFunc:EXTAtten
This command is used to specify the attenuation value as a multiplier to the given
scale factor on the specified channel. The channel is specified by x.
The query form of this command returns the user-specified attenuation.
Group
Vertical
Syntax
CH<x>:PROBEFunc:EXTAtten <NR3>
CH<x>:PROBEFunc:EXTAtten?
Related Commands
CH<x>:PROBEFunc:EXTDBatten
Arguments
<NR3> is the attenuation value, which is specified as a multiplier in the range
from 1.00E-10 to 1.00E+10.
Examples
CH1:PROBEFUNC:EXTATTEN 167.00E-3 specifies an external attenuation,
which is c onnected between the your input signal and the input of the probe
attached to Channel 1.
CH2:PROBEFUNC:EXTATTEN? might return :CH2:PROBEFUNC:EXTATTEN
1.0000E+00
, indicating that the probe attached to Channel 2 is connected
directly to the user's signal.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-201