Commands listed in alphabetical order
Examples
TRIGger:A:LOG
Ic:DELTatime 4.5e-9
sets the delta time to 4.5 ns.
TRIGger:A:LOGIc:DELTatime ? might return
:TRIGGER:A:L
OGIC:DELTATIME 4.0E-9
indicating the delta time is set to
4ns.
TRIGger:{A|B}:LOGIc:FUNCtion
This command sets or queries the logical combination of the input channels for
logic triggers. This command is equivalent to selecting Logic for the Trigger
Type, and setting or viewing the Define Logic.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:FUNCt ion {AND|NANd|NOR|OR}
TRIGger:{A|B}:LOGIc:FUNCt ion?
Arguments
AND specifies to trigger if a ll conditions are true.
NANd specifies to trigger if any of the c onditions are false.
NOR specifies to trigger if all conditions are false.
OR specifies to trigger if any of the conditions are true.
Examples
TRIGGER:A:LOGIC:FUNCTION AND sets the logical combination of channels to
be true when all conditions are true.
TRIGGER:A:LOGIC:FUNCTION? might return :TRIGGER:A:LOGIC:FUNCTION
NAND
, indicating that the instrument will trigger if the AND logic conditions
are false.
TRIGger:{A|B}:LOGIc:INPut:CLOCk:SOUrce
This command specifies the channel to use as the clock source for logic trigger.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:INPut :CLOCk:SOUrce {CH<x>|CH<x>_D <y>}
TRIGger:{A|B}:LOGIc:INPut :CLOCk:SOUrce?
2-852 MSO54, MSO56, MSO58, MSO58LP Programmer