Commands listed in alphabetical order
BUS:B<x>:ETH
ERnet:DATAPLUSTHRESHold
This command sets or queries the E thernet D+ source threshold for the specified
bus. This threshold only applies when the Ethernet signal type is single ended.
The bus is spe
cified by x.
Conditions
Requires option SR-ENET.
Group
Bus
Syntax
BUS:B<x>:ETHERnet:DATAPLU STHRESHold <NR3>
BUS:B<x>:ETHERnet:DATAPLU STHRESHold?
Related Commands
BUS:B<x>:ETHERNET:SOUrce:DMINus, BUS:B<x>:ETHERNET:SOUrce:
DPLUs, BUS:B<x>:ETHERnet:SIGNALTYpe
Arguments
<NR3> is the Ethernet D+ source threshold
for the specified bus.
Examples
BUS:B3:ETHERNET:DATAPLUST HRESHOLD 250.0E-3 sets the Bus 3 Ethernet
D+ source threshold to 250.0000 mV.
BUS:B2:ETHERNET:DATAMINUS THRESHOLD? might return
:BUS:B2:ETHERNET:DATAPLUS THRESHOLD 1.0 indicating the Bus
2 Ethernet D+ source threshold is set to 1.0 V.
BUS:B<x>:ETHERnet:IPVFOUR
This command sets or queries whether IPV4 packets are available for triggering
on Ethernet. The bus is specified by x.
Conditions
Requires option SR-ENET.
Group
Bus
Syntax
BUS:B<x>:ETHERnet:IPVFOUR {YES|NO}
BUS:B<x>:ETHERnet:IPVFOUR ?
Arguments
YES specifies that IPV4 packets are available.
NO specifies that IPV4 packets are not available.
2-134 MSO54, MSO56, MSO58, MSO58LP Programmer