Commands listed in alphabetical order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:GLOBa lref {OFF|ON|0|1}
MEASUrement:MEAS<x>:GLOBa lref?
Arguments
OFF allows ref levels to be set separately for each measurement.
ON applies the same ref levels to all measureme nts.
0 allows ref levels to be set separately for each measurement.
1 applies t he same ref levels to all measurements.
Examples
MEASUrement:MEAS1:GLOBalr ef 0 allows ref leve l s to be set sepa r ately for
each measurement.
MEASUrement:MEAS1:GLOBalr ef? might return
:MEASUREMENT:MEAS1:GLOBAL REF 1 indicating the same ref levels apply
to all measurements.
MEASUrement:MEAS<x>:HIGHREFVoltage
This command sets or queries the high reference voltage value for the 'time
outside leve l' measurement. Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:HIGHR EFVoltage <NR3>
MEASUrement:MEAS<x>:HIGHR EFVoltage?
Arguments
<NR3> is the high reference voltage value for the selected configuration.
Examples
MEASUrement:MEAS1:HIGHREF Voltage 1.5 sets the high reference voltage
to 1.5 V.
MEASUrement:MEAS1:HIGHREF Voltage? might return
:MEASUREMENT:MEAS1:HIGHRE FVOLTAGE 1.0000 indicating the
high reference voltage is set to 1.0 V.
2-456 MSO54, MSO56, MSO58, MSO58LP Programmer