Commands listed in alphabetical order
Examples
TRIGGER:A:LOW
ERTHRESHOLD:CH2 1.3
sets the A trigger threshold voltage
for Channel 2 to 1.3 V.
TRIGGER:A:LO
WERTHRESHOLD:CH3?
might return
:TRIGGER:A:LOWERTHRESHOLD :CH3 1.2000E+00, indicating that the A
trigger threshold voltage for Channel 3 is 1.2 V.
TRIGger:{A
|B}:PULSEWidth:HIGHLimit
This command species the upper limit to use, in seconds, when triggering on
detection of a pulse whose duration is inside or outside a range of two values.
(Use TRIG
ger:{A|B}:PULSEWidth:LOWLimit to specify the lower value of
the range.)
Group
Trigger
Syntax
TRIGger:{A|B}:PULSEWidth: HIGHLimit <NR3>
TRIGger:{A|B}:PULSEWidth: HIGHLimit?
Related Commands
TRIGger:{A|B}:PULSEWidth:WHEn, TRIGger:{A|B}:PULSEWidth:LOWLimit
Arguments
<NR3> is a oating point number that represents the higher value of the range.
Examples
TRIGger:A:PULSEWidth:HIGH Limit 110.0E-9 sets the high limit to 110.0
ns.
TRIGger:A:PULSEWidth:HIGH Limit? might return
:TRIGGER:A:PULSEWIDTH:HIG HLIMIT 178.88000E-9 indicates
the high limit is set to 178.88 ns.
TRIGger:{A|B}:PULSEWidth:LOGICQUALication
This command species whether or not to use logic qualication for a pulse
width trigger.
Group
Trigger
Syntax
TRIGger:{A|B}:PULSEWidth: LOGICQUALification {ON|OF F}
TRIGger:{A|B}:PULSEWidth: LOGICQUALification?
2-856 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals