Commands listed in alphabetical order
POWer:POWer<
x>:GATing
This command sets or queries the gating type for the specified power measurement
badge. The power measurement badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:GATing {NONE | CURSOR | SCREEN | LOGIC}
POWer:POWer<x>:GATing?
Arguments
NONE makes measurement across the entire waveform record.
CURSOR makes measurements on that portion of the waveform between the
cursors. Selecting Cursors opens cursors on the measurement source . Set the
cursors so that the waveform area of interest is in between the cursors.
SCREEN takes measurements on that portion of the waveform shown in the
display. When Zoom is on, the display is the zoom window.
LOGIC takes measurements only when the logical state of a specified waveform is
true.
Examples
POWer:POWer1:GATing CURSOR sets the gating type for the power
measurement badge 1 as Cursor.
POWer:POWer2:GATing? might return SCREEN indicating the configured
gating type for the power measurement badge 2.
POWer:POWer<x>:GATing:GLOBal
This command sets or queries the gating settings for the specified power
measurement badge. The power measurement badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:GATing:GLO Bal {ON|OFF1|0}
POWer:POWer<x>:GATing:GLO Bal?
2-536 MSO54, MSO56, MSO58, MSO58LP Programmer