Commands listed in alphabetical order
Related Commands
POWer:POWer<x
>:REFLevels:METHod, POWer:POWer<x>:REFLevels:
ABSolute:TYPE
Arguments
<NR1> ranges
from –40000 to 40000
Examples
POWer:POWer1:REFLevels:ABS olute:RISELow –1 sets the low reference
level for rising edge as –1 volt.
POWer:POWer<x>:REFLevels:ABSolute:RISEMid
This command sets or queries the rising edge for mid reference level in absolute
units for the specied power measurement badge. The power measurement badge
is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
P
OWer:POWer<x>:REFLevels:ABSo lute:RISEMid
<NR1>
POWer:POWer<x>:REFLevels:A BSolute:RISEMid?
Related Commands
POWer:POWer<x>:REFLevels:METHod, POWer:POWer<x>:REFLevels:
ABSolute:TYPE
Arguments
<NR1> ranges from –40000 to 40000
Examples
POWer:POWer1:REFLevels:ABS olute:RISEMid 0 sets the mid reference
level for rising edge as 0 volt.
POWer:POWer<x>:REFLevels:ABSolute:TYPE
This command sets or queries the type of measurement levels when reference
level is set to absolute for the specied power measurement badge. The power
measurement badge is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
MSO54, MSO56, MSO58, MSO58LP Programmer 2-555

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals