Commands listed in alphabetical order
MEASUrement:
MEAS<x>:GATing:HYSTeresis
This command sets or queries the gating hysteresis value when the gating type is
logic. Measurements are speci ed by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:GATin g:HYSTeresis <NR3>
MEASUrement:MEAS<x>:GATin g:HYSTeresis?
Arguments
<NR3> is the g ating hysteresis.
Examples
MEASUrement:MEAS1:GATing: HYSTeresis 25.0E-3 sets the hysteresis
to 25 mV.
MEASUrement:MEAS1:GATing: HYSTeresis? might return
:MEASUREMENT:MEAS1:GATING :HYSTERESIS 30.0000E-3 indicating the
hysteresis is set to 30.0 mV.
MEASUrement:MEAS<x>:GATing:LOGICSource
This command sets or queries the gating data source when the gating type is logic.
The measurement number is specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:GATin g:LOGICSource
{CH<x>|MATH<x>|REF<x>}
Arguments
Arguments are the sources for logic gating.
Examples
MEASUrement:MEAS1:GATing: LOGICSource CH3 sets the gating logic source
to channel 3.
MEASUrement:MEAS1:GATing: LOGICSource? might return
:MEASUREMENT:MEAS1:GATING :LOGICSOURCE CH2 indicating the gating
logic source is channel 2.
2-454 MSO54, MSO56, MSO58, MSO58LP Programmer