Commands listed in alphabetical order
Group
Measurement
Syntax
MEASUrement:REFLevels:ABS olute:TYPE {SAME|UNIQue}
MEASUrement:REFLevels:ABS olute:TYPE?
Arguments
SAME specifies that the absolute levels are set the same.
UNIQue specifies that the absolute levels can be set independently.
Examples
MEASUrement:REFLevels:ABS olute:TYPE UNIQUE specifies that the
absolute levels can be set independently.
MEASUrement:REFLevels:ABS olute:TYPE? might return
:MEASUREMENT:REFLEVELS:AB SOLUTE:TYPE SAME indicating the absolute
levels are set the same.
MEASUrement:REFLevels:BASETop
This command sets o r queries the method used to calculate the TOP and BASE,
used to calculate reference levels for the measurement.
Group
Measurement
Syntax
MEASUrement:REFLevels:BAS ETop
{AUTO|MINMax|MEANhistogra m|MODEhistogram|EYEhistogram
MEASUrement:REFLevels:BAS ETop?
Arguments
Argumentsarethebasetopmethods.
Examples
MEASUREMENT:REFLEVELS:BAS ETOP MINMAX selects the MINMAX base top
method.
MEASUREMENT:REFLEVELS:BAS ETOP? might return
:MEASUREMENT:REFLEVELS:BA SETOP MINMAX indicating the base top method
is MINMAX.
MEASUrement:REFLevels:JITTERMODE
This command sets or queries how often reference levels are calculated on Jitter
measurements. If the mode is set to Latch, ref levels are calculated only on the
first acquisition after a statistics reset. If it is set to Continuous, reference levels
are calculated on every acquisition.
2-500 MSO54, MSO56, MSO58, MSO58LP Programmer