Commands listed in alphabetical order
MEASUrement:
CLOCKRecovery:ADVanced:METHod
This command sets or queries the global advanced clock
recovery method. This will affect measurements whose
:MEASUremen
t:MEAS<x>:CLOCKRecovery:GLOBal flag is set to
1.
Group
Measurement
Syntax
MEASUrement:CLOCKRecovery :ADVanced:METHod
{NONE|NOMinal|PATTern}
MEASUrement:CLOCKRecovery :ADVanced:METHod?
Arguments
NONE sets to use no advanced CRD method.
NOMinal sets the advanced CRD method to Nominal Data Rate.
PATTern sets the advanced CDR meth
od to use a Known Data P attern.
Examples
MEASUREMENT:CLOCKRECOVERY :ADVANCED:METHOD NOMINAL sets the CDR
method to use a Known Data Pattern.
MEASUREMENT:CLOCKRECOVERY :ADVANCED:METHOD? might return
:MEASUREMENT:CLOCKRECOVER Y:ADVANCED:METHOD NONE indicating that no
advanced CRD method will be used.
MEASUrement:CLOCKRecovery:CLOCKFrequency
This command sets or queries the global clock frequency used when fixed constant
clock recovery is used for the measurement. This will affect measurements whose
:MEASUrement:MEAS<x>:CLOCKRecovery:GLOBal flag is set to 1.
Group
Measurement
Syntax
MEASUrement:CLOCKRecovery :CLOCKFrequency <NR3>
MEASUrement:CLOCKRecovery :CLOCKFrequency?
Arguments
<NR3> is the global clock frequency used with Constant Clock - Fixed clock
recovery method.
Examples
MEASUREMENT:CLOCKRECOVERY :CLOCKFREQUENCY 2.0E+9 sets the clock
frequency to 2.0 GHz.
2-392 MSO54, MSO56, MSO58, MSO58LP Programmer