Commands listed in alphabetical order
VERTical:DES
KEW:TO:SOUrce
This command sets or queries the target channel for performing channel-to-channel
deskew adjustment. Target sources can be any of the live analog channels.
Group
Vertical
Syntax
VERTical:DESKEW:TO:SOUrce CH<x>
VERTical:DESKEW:TO:SOUrce?
Arguments
Arguments are the live analog channels.
Examples
VERTICAL:DESKEW:TO:SOURCE CH4 sets the deskew to source is channel 4.
VERTICAL:DESKEW:TO:SOURCE? might return
:VERTICAL:DESKEW:TO:SOURCE CH2 indicating the deskew to source is
channel 2.
*WAI (No Query Form)
The *WAI (Wait) command (no query form) prevents the instrument from
executing further commands or queries until all pending commands that generate
an OPC message are complete. This command allows you to synchronize the
operation of the instrument with your application program. For more information,
refer to Synchronization Methods.
Group
Status and Error
Syntax
*WAI
Related Commands
BUSY?, *OPC
Examples
*WAI prevents the instrument from executing any further commands or queries
until all pending commands that generate an OPC message are complete.
WAVFrm? (Query Only)
This query-only command provides the Tektronix st
andard waveform query
which returns the waveform preamble followed by the waveform data for the
MSO54, MSO56, MSO58, MSO58LP Programmer 2-889

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals