Commands listed in alphabetical order
Arguments
<NR1> is the hor
izontal record length.
Examples
HORIZONTAL:RECORDLENGTH 1000 setstherecordlengthto1000samples.
HORIZONTAL:RECORDLENGTH? might return : HORIZONTAL:RECORDLENGTH
1000000
, indicating that the record length is set to 1,000,000 samples
HORizontal:ROLL? (Query Only)
Queries the horizontal roll mode status.
Group
Horizontal
Syntax
HORizontal:ROLL?
Returns
ON indicates roll mode is active.
OFF indicates that summary frame is off.
<NR1> a 0 indicates roll mode is off; any other v alue activates roll mode.
Examples
HORIZONTAL:ROLL? might return :HORIZONTAL:ROLL 0 indicating roll mode
is off.
HORizontal:SAMPLERate
This command sets or queries the horizontal sample rate.
Group
Horizontal
Syntax
HORizontal:SAMPLERate <NR3>
HORizontal:SAMPLERate?
Arguments
<NR3> is the horizontal sample rate in samples per second.
Examples
HORIZONTAL:SAMPLERATE 1e6 sets the sample rate to 1 million samples per
second.
HORIZONTAL:SAMPLERATE? might return :HORIZONTAL:SAM PLERATE
5.0000E+6
, indicating that the sample rate is set to 5 million samples per second.
2-354 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals