Commands listed in alphabetical order
TRIGger:B:RE
SET:TYPe
This command sets or queries the type of A→B sequential t rigger reset. If the B
trigger reset is active, the reset criteria are part of the B triggering sequence. If
the reset con
ditions defined by the reset type are met, the instrument must start
over searching for a new occurrence of the A event.
You must ide
ntify a trigger Source and Threshold for each reset type, except
for the Timeout trigger type.
NOTE. If a reset condition occurs, the reset criteria itself is reset and must start
over.
Group
Trigger
Syntax
TRIGger:B:RESET:TYPe {NONe|TIMEOut| EDGE}
TRIGger:B:RESET:TYPe?
Arguments
NONe defeats the trigger reset feature.
TIMEOut initiates a reset if the timeout conditions specified by
TRIGger:B:RESET:TIMEOut are met.
EDGE initiates a reset if the edge trigger conditions are met.
Examples
TRIGGER:B:RESET:TYPE NONE deactivates the sequential trigger reset.
TRIGGER:B:RESET:TYPE? This query might return :TRIGGER:B: RESET:TYPe
TIMEOUT
, indicating that the sequential trigger reset is active following a timeout.
TRIGger:B:STATE
This command sets o r queries the state of B trigger activity. If the B trigger state
is on, the B trigger is part of the triggering sequence. If the B trigger state is off,
then only the A trigger causes the trigger event.
Group
Trigger
Syntax
TRIGger:B:STATE {ON|OFF|<NR1>}
TRIGger:B:STATE?
Related Commands
TRIGger:A:MODe
2-882 MSO54, MSO56, MSO58, MSO58LP Programmer