Commands listed in alphabetical order
NOTEQual argum
ent sets the oscilloscope to trigger if a runt pulse is d etected with
width greater than or less than (but not equal to) the time period specied in
TRIGger:{A|B}:RUNT:WIDth within a ±5% tolerance.
Examples
TRIGger:A:RUNT:WHEn MORETHAN sets the runt trigger to occur when the
oscilloscope detects a runt in a pulse wider than the specied width.
TRIGger:A:RUNT:WHEn? might return :TRIGGER:A:PULSE:RUNT:WHEN
OCCURS
indicating that a runt trigger will occur if the oscilloscope detects a runt
of any detectable width.
TRIGger:{A|B}:RUNT:WIDth
This command species the width, in seconds, for a runt trigger.
Group
Trigger
Syntax
TRIGger:{A|B}:RUNT:WIDth <NR3>
TRIGger:{A|B}:RUNT:WIDth?
Related Commands
TRIGger:{A|B}:RUNT:WHEn
Arguments
<NR3> is a oating point number that species the minimum width, in seconds.
Examples
TRIGger:A:RUNT:WIDth 15E-6 sets the m inimum width o f the pulse runt
trigger to 15 µs.
TRIGger:A:RUNT:WIDth? might return :TRIGGER:A:PULSE:RUNT:WIDT H
2.0000E-09
indicating that the minimum width of a pulse runt trigger is 2 ns.
TRIGger:{A|B}:SETHold:CLOCk:EDGE
This command species the clock edge polarity for setup and hold triggering.
Group
Trigger
Syntax
TRIGger:{A|B}:SETHold:CLO Ck:EDGE {FALL|RISe}
TRIGger:{A|B}:SETHold:CLO Ck:EDGE?
2-862 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO56 and is the answer not in the manual?

Tektronix MSO56 Specifications

General IconGeneral
BrandTektronix
ModelMSO56
CategoryTest Equipment
LanguageEnglish

Related product manuals