Syntax
MEASUrement:M
EAS<x>:GATing:ACTive {HIGH|LOW}
MEASUrement:MEAS<x>:GATing :ACTive?
Arguments
HIGH takes a measurement when logic gating is High.
LOW takes a measurement when logic gating Low.
Examples
MEASUrement:MEAS1:GATing:A CTive LOW specifies taking measurements
when logic gating is Low.
MEASUrement:MEAS1:GATing:A CTive? might return
:MEASUREMENT:MEAS1:GATING: ACTIVE HIGH indicating measurements are
taken when logic gating is High.
MEASUrement:MEAS<x>:GATing:GLOBal
This command sets or queries the gating settings global flag. Measurements are
spec
ified by x.
Group
Measurement
Syntax
M
EASUrement:MEAS<x>:GATing:GL OBal {OFF|ON|0|1}
MEASUrement:MEAS<x>:GATing :GLOBal?
Arguments
OFF specifies gate settings can be changed independently for each individual
measurement.
ON applies global gate settings to all the measurements' gate settings.
0 specifies gate settings can be changed independently for each individual
measurement.
1 applies global gate settings to all the measurements' gate settings.
Examples
MEASUrement:MEAS1:GATing:G LOBal OFF specifies gate settings can be
changed independently for each individual measurement.
MEASUrement:MEAS1:GATing:G LOBal? might return
:MEASUREMENT:MEAS1:GATING: GLOBAL 1 indicating that gating
settings apply to a ll measurements.