Commands listed in alphabetical order
Syntax
TRIGger:{A|B}
:BUS:B<x>:LIN:IDentifier:VAL ue <QString>
TRIGger:{A|B}:BUS:B<x>:LI N:IDentifier:VALue?
Arguments
<QString> is the binary address string used for LIN trigger if the trigger
condition is ID or IDANDDATA.
Examples
TRIGGER:A:BUS:B1:LIN:IDEN TIFIER:VALUE "110010" sets the identifier
value to 110010.
TRIGGER:A:BUS:B1:LIN:IDEN TIFIER:VALUE? might return
TRIGGER:A:BUS:B1:LIN:IDEN TIFIER:VALUE "XXXXXX" indicating the
identifier value is XXXXXX.
TRIGger:{A|B}:BUS:B<x>:MIL1553B:COMMAND:ADDRess:HIVALue
This command sets or queries the high value when triggering on command word
addresses for a MIL-STD-1553 bus. The bus number is specified by x. The
trigger condition must be set to COMMAND, and the address qualifier must be
INrange or OUTrange.
Conditions
Requires the SR-AERO Triggering and Analysis application
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:MI L1553B:COMMAND:ADDRess:HIVALue
<QString>
TRIGger:{A|B}:BUS:B<x>:MI L1553B:COMMAND:ADDRess:HIVALue?
Arguments
<QString> is the address value.
Examples
TRIGGER:A:BUS:B1:MIL1553B :COMMAND:ADDRESS:HIVALUE "X1000"
sets the value to X1000.
TRIGGER:A:BUS:B1:MIL1553B :COMMAND:ADDRESS:HIVALUE? might return
"XXXXX", indicating that the value is XXXXX.
TRIGger:{A|B}:
BUS:B<x>:MIL1553B:COMMAND:ADDRess:QUALifier
This command sets or queries the qualifier to be used when triggering on
command word addresses for a MIL-STD-1553 bus. The bus number is specified
by x. The trigge
r condition must be set to COMMAND.
2-814 MSO54, MSO56, MSO58, MSO58LP Programmer