Commands listed in alphabetical order
Syntax
POWer:POWer<x
>:SWITCHINGLOSS:VLEVELPct
<NR1>
POWer:POWer<x>:SWITCHINGLO SS:VLEVELPct?
Related Commands
POWer:POWer<x>:SWITCHINGLOSS:LEVELUNIts
Arguments
<NR1> ranges from 0.0001 to 90
Examples
POWer:POWer1:SWITCHINGLOSS :VLEVELPct 7 sets the voltage level
(Ton-Start & Stop) value as 7 percentage for the switching loss measurement in
the power measurement badge 1.
POWer:POWer<x>:SWITCHINGLOSS:VSOURce
This comma nd sets or queries the voltage source for the switching loss
measurement in the specified power measurement badge. The power measurement
badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:SWITCHINGLO SS:VSOURce {CH<x> | MATH<x> | REF<x>}
POWer:POWer<x>:SWITCHINGLO SS:VSOURce?
Arguments
CH<x> = A channel specifier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x> = A math waveform specifier; <x> is ≥1.
REF<x> = A reference waveform specifier; <x> is ≥1.
Examples
POWer:POWer1:SWITCHINGLOSS :VSOURce REF1 sets the voltage source
for the switching loss measurement in the specified power measurement badge
as REF1.
POWer:POWer2:SWITCHINGLOSS :VSOURce? might return REF3 indicating the
voltage source for switching loss measurement of Power 2 power measurement
badge.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-593