EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 102
F- 9.0.24 ISOLATE FLOW PUMP INPUT FLOW
a) Obtain a 1000 ml (or larger) graduated cylinder!
b) Service Mode’s Flow Pressure is used to isolate the Flow Pump’s input circuit!
DO NOT follow the screen instructions! Perform no adjustments until instructed!
c) Enter Service Mode → Calibrate Hydraulics → Flow Pressure.
d) Press ‘EnterTHEN ENSURE the FLOW MOTOR shaft
is rotating!
e)
NOTE the gauge reading for later.
f) Figure right, remove the Flow Pump’s INPUT (clear)
tubing and measure flow from it for one (1) minute.
g) Press ‘Enter’ to return to “Calibrate Hydraulics”.
h) Reattach the Flow Pump’s tubing!
i) TWO (2) possible scenarios:
1) IF (and ONLY if) MORE THAN 800 ml collected! See procedure number F- 9.0.25 (page 102).
2) IF LESS THAN 800 ml collected! Proceed to page 105, procedure number F- 9.0.4.
F- 9.0.25 INPUT FLOW PUMP MORE THAN 800 ML
a) Place the machine into RINSE!
b) As noted above, was pressure between 35 and 36 psi?
Yes Between 35 and 36 psi! Good static pressure but NOT dynamic! TWO (2) possible bad
components: 1) Bad Flow Pump head OR; 2) Bad Flow Pump motor (possibly brushes).
No Less than 35 psi! See procedure number F- 9.0.26 (page 103).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals