2008T Troubleshooting Guide
P/N 490292 Rev. A
Scenario #3: Referring to Figure C, for a ONE MINUTE PERIOD if EVER a
continuous weak stream that LASTS for five (5) seconds or
longer: ENSURING the drain tube OPENING was pointed up, proceed to
page 66, procedure number F- 6.2.0.
Figure C
Scenario #4: IF No Flow EVER: If (and ONLY if) the [Dialysate Flow] window says “OFF” OR
is blinking return to (ABOVE) procedure number F- 5.2.0
(
page 57). If [Dialysate Flow] does NOT say “OFF” and is
not blinking proceed to
page 67, procedure number
F- 6.3.0.
F- 6.1.0 PULSE – STOP – PULSE OR ‘DRIBBLE’ / ISOLATE FOR STUCK OPEN BC VALVE
A) Reattach the drain tubing to the station!
B) ENSURING [Dialysate Flow] is on ENSURE a good drain connection by listening for or visually
verifying flow.
C) If a TMP alarm is NOT present skip to part D. If a TMP alarm is present: 1) Press and release the
‘Reset’ key then immediately press and hold it for three (3) seconds; 2) Allow thirty (30) seconds;
3) If a TMP alarm reoccurs attempt RESET up to twice more BEFORE continuing to part D.
D) Allow thirty (30) seconds BEFORE continuing to part E!
E) Call debug screen 4 to watch PDIA (left column) AND ADIA (right
column) for TWO (2) minutes. BOTH should REMAIN between
1.5 and 8.5 AND STABLE i.e. does NOT change more than 0.3
per minute!
F) TWO (2) possible scenarios:
1) IF (and ONLY if) PDIA AND ADIA is REMAINING between 1.5 and 8.5 AND is STABLE:
Proceed to page 62, procedure number F- 6.1.4.
2) IF PDIA and / or ADIA is
NOT remaining between 1.5 and 8.5 and / or is NOT stable: See
procedure number F- 6.1.1 (page 60).