2008T Troubleshooting Guide
P/N 490292 Rev. A
Yes Cables were OKAY! See procedure number FIL- 1.0.8 (page 458).
No Problem with a cable! After the repair, perform parts a THROUGH d below:
a) Turn the machine OFF then allow seven (7) minutes BEFORE continuing to part b.
b) Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter’)!
c) Allow [Conductivity] to increase to more than 13.0 mS.
d) Call debug screen 1.Watch FILACT for two (2) minutes. If REMAINING = 0 the problem
is solved. If still going ever = 1 return to (ABOVE) procedure number FIL- 1.0.3 (page
456).
FIL- 1.0.8 OLC ACTIVATED?
From debug screen 0, once again, does the OLC POST Dialyzer Cond symbol appear?
Yes POST Dialyzer Cond symbol appears! A problem with Post Dialyzer Cond. Close the shunt
door THEN proceed to page 417, procedure number CO- 10.0.10.
No The OLC option will have to be activated as a bad OLC reading may cause Filling Programs EVEN
IF the OLC option is not activated! Perform parts a THROUGH i below:
a) Enter Service Mode → Options → Hardware Options.
b) Next to OLC (upper right) place the ‘X’ in the “Yes” box and press the
keyboard’s ‘Enter’ key. The ‘X’ turns blue.
c) Turn the machine off then bacl on.
d) Return to Dialysis Program (“Select Program”→ ‘Dialysis’ → ‘Enter’)!
e) Call debug screen 0 and confirm the Post Dialyzer OLC symbols now appear.
f)
The shunt door MUST be open!
g) From the Home screen, ENSURING [Dialysate Flow] is set at 800 ml/min, allow Conductivity
to reach between 13.0 and 14.5 mS!
h) Allow two (2) additional minutes for Post Conductivity to stabilize!
i) Call debug screen 0. Are PRE AND POST Cond within 0.4 of each other now?
Yes PRE AND POST Conductivity within 0.4 mS of each other! See procedure
number FIL- 2.0.0 (page 459).
No PRE AND POST Conductivity ARE NOT within 0.4 mS of each other! A problem with
Post Dialyzer Cond. Close the shunt door and proceed to page 417, procedure
number
CO- 10.0.10 (page 417).