2008T Troubleshooting Guide
P/N 490292 Rev. A
SECTION 21 – HYDRAULIC LEAKS
NOTE! These procedures do NOT troubleshoot Vent Tubing overflow!
A) Turn the machine OFF!
B) ENSURE the ‘Incoming Water’ and ‘To Drain’ tubing (Figure right)
are not reversed connected!
C) ENSURE the ‘To Drain’ tubing is not kinked AND, if using a drain
’Quick Connector’, it is attached PROPERLY to the station drain!
D) Reattach any ‘blown off’ tubing and dry the area!
E) Until
INSTRUCTED OTHERWISE, direct the Fluid Sample
Connector (Figure right) FIGURE104into a bucket!
F) Per Figure 104 below, was tubing ‘blowing off’ a UF Check
Valve?
Yes UF Check Valve tubing ‘blowing off’! See procedure
number LEAKING- 1.0.0 (page 599).
No UF Check Valves okay! Proceed to
page 600, procedure
number LEAKING- 1.1.0.
HYDRAULICS TOP VIEW
Figure 104 – UF Check Valves #63 AND #64