2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 744
SECTION 31 – VALVE 104 OR VALVE 105 ERROR
A) Call debug screen 15. THREE (3) possible scenarios 1) or 2) or 3) below:
1) IF (and ONLY if) V104 Er (3
rd
column from left) = 1: Referring to the Figure
below, there is a problem with Valve #104 and / or Valve #108. See part B.
2) IF (and ONLY if) V105 Er (2
nd
column from left) = 1: Referring to the Figure
below, there is a problem with Valve #105. See part B.
3) IF BOTH V105 Er AND V104 Er = 0: You are in the wrong Section! Neither valve is reporting a
problem!
B)
Turn the machine OFF!
C) Set your CALIBRATED volt meter to RESISTANCE (Ω)!
D) TWO (2) possible scenarios:
1) IF (and ONLY if) V104 Er = 1: Problem with Valve #104 AND / OR Valve 108! See procedure
number V- 1.0.0 (page 744).
2) IF V105 Er = 1: Problem with Valve #105! Proceed to
page 745, procedure number V- 2.0.0.
Hydraulics Front View, Valves #105, #104 and #108 Locations
V- 1.0.0 V104 Er = 1 / PROBLEM WITH VALVE #104 AND / OR VALVE #108
A) ENSURE Valve #104’s and #108 wires terminate properly at their solenoid
terminals (Figure right). If not, this is may be the problem!
B)
Unplug both Valve #104’s wires from its solenoid!
Parts C through E continued on next page

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals