2008T Troubleshooting Guide
P/N 490292 Rev. A
F- 1.0.5 RESET MACHINE / DEFEAT IDLE FLOW / CHECK PUMPS
A) Turn the machine OFF!
B) Open the shunt door and leave it OPEN till instructed!
C) Turn the machine on and return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter’)!
D) From the Home screen, press the [Dialysate Flow] window.
E) Set [Dialysate Flow] to 800 ml/min and press ‘Enter’.
F) Call debug screen 1. If FILACT = 0 continue to part G. If (and ONLY if) FILACT = 1 return to (ABOVE)
procedure number F- 1.0.4 (page 28)
G)
Per the Figure below, check the Flow AND Deaeration Motor shafts for counterclockwise (CCW)
rotation. BOTH rotating CCW?
Yes BOTH motors rotating! See procedure number F- 1.0.6 (page 31).
No A motor is NOT rotating CCW. Perform parts a AND b below:
a) Close the shunt door!
b) NOTING which motor is not rotating properly, proceed to page 141,
TROUBLESHOOTING MOTORS
Figure 7 – Deaeration AND Flow Motor