EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 111
F- 10.0.0 BAD ACFS SIGNAL
a) Figure right, ENSURE the CFS #10
Pressure Transducer’s connector is
plugged PROPERLY into the
distribution board.
b) See procedure number F- 10.1.0
(page 111).
F- 10.1.0 CFS PLUGGED IN PROPERLY
a) Call debug screen 10. If all procedures were performed correctly PDIA (middle column)
= between 3.5 and 6.5.
b) Unplug the CFS’s connector from position X10, “CFS”.
c) Using a flashlight, check inside the vacant “X10position. If corrosion or damaged pins are located
this may be the problem!
d) Figure right, plug the Dialysate
Pressure Sensor’s (#9) connector
into the CFS Pressure Sensor’s. “P-
DIAL” “CFS”.
e) Does
ACFS (lower
right) go to between 4.0
and 6.0?
Yes ACFS between 4.0 and 6.0! The CFS Pressure Sensor #10* is bad. *To LOCATE the
CFS Pressure Sensor #10 refer to Figure 21 (page 100).
No ACFS is less than 4.0 OR more than 6.0! Leaving PDIA in CFS for now, FIVE (5) possible bad
components: 1) Bad Actuator-Test Board
a
OR; 2) Bad Sensor Board
b
OR; 3) Bad Sensor
Board cable OR; 4) Bad distribution board OR;5) Bad motherboard.
a
A) With the power off, swap in a known good Actuator-Test
1
Board; B) IMPORTANT!
Return Dialysis Program (“Select Program” → ‘Dialysis’ → Enter’); C) If ACFS is between
4.0 and 6.0 the previous Actuator-Test Board is bad.
1
To LOCATE the board refer to Figure 4A (page 10).
b
A)
With the power off, swap in a known good Sensor Board
2
; B) Place the machine into
T and C Mode (refer to OPERATING MODES (page 19); C) IMPORTANT! Return to
Dialysis Program)!; D) If ACFS now = between 4.0 and 6.0 the previous Sensor Board is
bad.
2
To LOCATE the Sensor board see Figure 4A (page 10).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals