EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #394 background imageLoading...
Page #394 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 390
CO- 6.6.0 POST DIALYZER COND / TROUBLESHOOTING POST DIALYZER COND
a) Recheck Loading Pressure (gauge in Rinse port). Unless there is an unannounced “No Water” or a
Flow Error alarm pressure cycles, to about 25 psi about every three (3) seconds! Do NOT continue
UNLESS cycling!
b) ENSURING the shunt door is CLOSED, if the external flow indicator‘s ‘bobis NOT moving up and
down there may be an unannounced No Water” or Flow Error. Do NOT continue UNLESS ‘bob’ is
moving up and down!
c) Check for air bubbles flowing through the dialyzer AND
the dialyzer lines. Air does NOT allow conductivity to
stabilize!
d) Figure right, using a flashlight, and WITHOUT
LOOKING AWAY for ONE (1) MINUTE watch for air
bubbles through the tubing at Cond Cell #7.
e) DO NOT continue UNLESS ABSOLUTELY SURE
there are NO AIR leaks!
f) See procedure number CO- 6.6.2 (page 390).
CO- 6.6.2 ISOLATE THE DIALYZER
a) Remove the dialyzer and reinstall the external meter to measure conductivity.
b) Close the shunt door then return to Service Mode → Calibrate Sensors → Cond Cells.
c) Follow the screen’s instructions through: 4. Waiting until the Conductivity value is stable.”
d) ENSURING the external flow indicator’s ‘bob’ is moving up and down, ALLOW six (6) FULL minutes to
ENSURE measured conductivity is between 13.0 and 14.4 mS.
e) Enter the measured conductivity into the [Conductivity] box and press ‘Enter’ twice.
f) This time do NOT install a dialyzer!
g) Press ‘Enter’. Allow up to twenty-five (25) FULL minutes OR until if an alarm occurs. If all FOUR ‘OLC
points’ advance successfully the screen says: “Press ‘CONFIRM’ to complete the calibration?”
Yes “Press ’CONFIRM…” appears! There may be a problem with the original dialyzer! ‘Escape’
the calibration. This time using a DIFFERENT PRIMED dialyzer return to
page 384,
procedure number CO- 6.2.0.
No Press ‘Confirm…” does NOT appear! TWO (2) possible scenarios:
1) IF (and ONLY if) “COND LESS THAN 10.0” OR “COND LESS THAN 15.0”
banner occurred: See (ABOVE) procedure number CO- 6.5.0 (page 389).
2) All OTHERS: A possible problem with the Cond Cell #7 Circuit. Proceed to
page 395, procedure number CO- 7.0.0.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals