2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 119
F- 13.0.0 ‘BOB’ IS MOVING WHEN IT’S NOT SUPPOSED TO
The Temperature AND / OR the Conductivity window is REMAINING RED BUT the external flow
indicator’s bob is moving!
a) Per the Figure BELOW, trace the wires from Valve #24’s AND #26’s distribution board connectors, to
Valve #24 and #26 in the hydraulics to ENSURE they are wired to the CORRECT valve!
b) Figure left, open the plastic female cover from
Valve #26’s distribution board connector.
c) Valve #26 wires MUST be soldered between pins two
(second from the top) AND five (bottom). If between the
top and bottom pins Valve #26 is probably connected to
distribution board position “V24and vice versa!
d) Return Valve #26’s connector to distribution board
position “V26”.
e) ENSURING [Dialysate Flow] is set to 800 ml/min AND is not blinking is the external flow indicator’s
‘bob’ still rising?
Yes ‘Bob’ moving! TWO (2) possible bad components: 1) Actuator-Test Board OR; 2) Valve
#24.
No ‘Bob’ NOT moving! If a Flow Error is still occurring, return to (ABOVE) procedure number
F- 6.0.0 (page 58). If the Flow Error never occurs do not continue.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals