2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 133
F- 16.0.5 LESS THAN 130 ML
Figure right, TWO (2) possible scenarios:
1) IF (and ONLY if) NOT equipped with a
bibag Connector: See procedure
number F- 16.0.7 (page 134).
2) IF equipped with a bibag Connector: See
parts a THROUGH e below:
a)
Per the Figure below, clamp the tubing attached at the top Valve #101.
b) With the plunger pushed fully in, once again, attach the syringe to the Deaeration Pump’s CLEAR
tubing.
c) Pull the plunger to the end of the barrel then remove the syringe.
d) MEASURE flow from the tubing looking for more than 130 ml every minute!
e) Repeat parts b through e up to
TEN (10) TIMES . Do you EVER get than 130 ml every minute?
Yes IF SURE of more than 130 ml per minute! Remove the clamp from Valve #101! THREE
(3) possible problems: 1) Loose bibag Interface Board ribbon cable (refer to Figure 4C,
page 11)) OR; 2) Bad Valve #101 OR; 3) Bad bibag Interface Board.
No After ten (10) attempts you DO NOT measure more than 130 ml per minute! Remove the
clamp from Valve #101 then see procedure number F- 16.0.7 (page 134).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals