EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #68 background imageLoading...
Page #68 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 64
F- 6.1.6 ISOLATE DIALYSATE PRESSURE (PDIAL)
Call debug screen 4 to again watch PDIA for TWO (2) FULL minutes. TWO (2) possible scenarios:
1) IF (and ONLY if) PDIA increases to more than 8.0 OR drops to less than 2.0: Return to
page 60, procedure number F- 6.1.1.
2) IF PDIA NEVER increases to more than 8.0 OR drops to less than 2.0: As NOTED in procedure
number F- 6.0.0 (page 58), drain flow was either: 1) Pulse Stop Pulse Stopetc. (NO ‘dribble’)
OR; 2) Pulse Stop Pulse ‘dribble’ (a noticeably weaker stream) etc. TWO (2) possible scenarios
i) or ii) below:
i) IF (and ONLY if) drain flow = Pulse – Stop – Pulse – Stop (NO ‘dribble’)
: See procedure
number F- 6.1.7 (page 64).
ii) IF drain flow = Pulse – Stop – Pulse ‘Dribble(two second slow stream): A procedure, in a
different Section, is performed next. NOTE this page and procedure number (F- 6.1.6), because
you may prompted to return to here, THEN perform parts A through C below:
A) BEFORE continuing to part b proceed to
page 580, to perform SECTION 19 - TESTING FOR
A LEAKING BALANCING CHAMBER DIAPHRAGM.
B) If a balancing chamber diaphragm leak was not located in part a, return to Dialysis Program
(“Select Program” → ‘Dialysis’ → Enter).
C) See
part b of procedure number F- 6.1.7 (page 64).
F- 6.1.7 ISOLATE FLOW PUMP CONTROL (FLWP)
a) If (and ONLY if) you have EVER seen [Temperature] ‘bounce’ up and down, more than 0.5º C, with
less than one (1) second in between, TWO (2) procedures: #1) NOTE this page and procedure number
(F- 6.1.7) for as you may be prompted to return here; #2) BEFORE continuing to part b proceed to
page 580, SECTION 19 - TESTING FOR A LEAKING BALANCING CHAMBER DIAPHRAGM.
b) Call debug screen 1. FLWP (upper right column) is a digital-to-analog value that controls Flow Pump
speed. It should remain stable (+/- 2). Without looking away watch FLWP for up to THREE (3) minutes.
Does it increase steadily more than 4, until a Flow Error occurs then falls only to increase again until a
Flow Error occurs, etc.?
Yes FLWP ramping up and down more than +/- 2: Possibly a tubing restriction somewhere. Start with
the Diasafe
®
filter possibly even inside the
housing.
No FLWP remaining stable: See procedure number F- 6.1.8 (page 65).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals