2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 94
F- 7.7.20 PART 3 GOOD FLOW / PART 4 LESS THAN 800 ML / RESTRICTED VALVES #38 AND #34
Balancing Chamber Valve #38 and/or Valve #34 is not opening. Also, the Actuator Cable or the distribution
board may be bad. These procedures isolate between these components:
a) Turn the machine OFF.
b) The Flow Pump remains plugged into the Deaeration Pump’s distribution board position.
c)
Per the Figure below, replace Valve #38 with a known good.
d) Plug the new valve #38’s connector into valve #31’s distribution board position, “V31”.
e) ENSURE the bucket with the RED dialyzer connector submersed in it is full!
f)
Turn the machine on but DO NOT press any other keys.
g) With “Select Program” up, measure drain flow, into the cylinder, for one (1) minute. More than 800 ml
every minute?
Yes More than 800 ml! Problem located! The previous valve #38 is bad!
No Less than 800 ml collected! See procedure number F- 7.7.22 (page 95).
.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals