EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #327 background imageLoading...
Page #327 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 323
T- 7.2.1 TEMPERATURE CONTROL
a) ENSURE the concentrate connectors are attached to acid and LIQUID bicarb.
b) From Service Mode → Calibrate Sensors → Temp Control.
c) ‘Sharply’ press ‘Enter’. The screen should say “3. Connect
an external temperature meter to the dialysate lines”.
d) Install a Temperature (°C) meter into the dialysate lines.
Figure right, flow MUST be bottom to top.
e) CLOSE THE SHUNT DOOR!
f) ‘Sharply’ press ‘Enter’. The screen says “5. Adjust the TEMP
DAC…” and the screen’s [TEMP DAC] data box (Figure right)
turns pale yellow / white.
g) Based on the [TEMP DAC] data box, TWO (2) possible scenarios:
1) IF (and ONLY if) between 120 and 180: See procedure number T- 7.2.2 (page 324).
2) IF NOT between 120 and 180: Perform parts a THROUGH d below:
a) Press [TEMP DAC], it turns bright yellow.
b) Set [TEMP DAC] to “140”.
c) ‘Sharply’ press ‘Enter’ ONCE. [TEMP DAC] MUST be pale yellow/white (NOT gray).
d) See procedure number T- 7.2.2 (page 324).
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals