2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 524
h) When “Get Pos TMP” momentarily appears ATMP normally drop to between +300 and +500.
i) When “Pos Flow Off” appears a twenty (20) second stabilization period begins. ATMP should settle
between +180 and +350.
NOTE: If FILACT EVER = 1, even if only once, during “Pos Flow OffNOTE this!
NOTE: If ATMP does not remain between +180 and +350Fail Pos Stabilize” will be listed in the Test
& Options screen possibly indicating an external fluid leak OR FILACT went to ‘1’.
j) If ATMP remained between +180 and +350 in part i a thirty (30) second HOLD period begins. “Pos
Flow Off” remains up AND ATMP should not change more than +/- 30
NOTE: If ATMP changes more than +/- 30 “Fail Pos Flow Off” will be listed in the Test & Options
screen possibly indicating a small external leak.
k) When the “Testing” banner appears IGNORE ATMP and FILACT. Do BOTH pass?
Yes BOTH pass! Proceed to
page 527, procedure number PHT- 4.0.0.
No Pressure Test(s) Failed! See procedure number
PHT- 3.1.2 (page 524).
PHT- 3.1.2 PRESSURE TEST(S) FAILED AGAIN
Press the Test & Options tab. FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1) IF (and ONLY if) Fail Remove Air” is listed: Proceed to
page 452, SECTION 8 FILLING
PROGRAM PROBLEMS
2) IF (and ONLY if) “Fail Neg Stabilizeis listed. TWO (2) possible scenarios i) or ii) below:
i) IF FILACT EVER = 1 during the repeat 30 second STABILIZATION Neg Flow Onperiod:
Proceed to
page 452, SECTION 8 FILLING PROGRAM PROBLEMS
ii) IF FILACT ALWAYS = 0 during the repeat 30 second “Neg Flow On” STABILIZATION
period: Proceed to
page 483, SECTION 9 TMP PROBLEMS!
3) IF (and ONLY if) “Fail Neg Flow Onis listed: THREE (3) possible scenarios i) or ii) or iii) below:
i) IF ATMP increased to more than -440: See procedure number PHT- 3.1.3 (page 525).
ii) IF FILACT EVER = 1 during the repeat 30 second HOLD Neg Flow On” period: Proceed
to
page 452, SECTION 8 FILLING PROGRAM PROBLEMS
iii) IF FILACT ALWAYS = 0 during the repeat 30 second “Neg Flow On” STABILIZATION
period: Proceed to
page 483, SECTION 9 TMP PROBLEMS
4) ALL Positive Test Failure listings i.e. Fail Get Pos TMP, Fail Pos Stabilize”, “Fail Pos Flow
Off”: Negative Pressure Test failures have priority over Positive Test failures. If any Negative Pressure
test failed see above Scenarios 2 or 3. If a Negative test DID NOT fail, TWO (2) possible scenarios i) or
ii) next page:

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals