2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 551
Yes BOTH pressure Tests pass! Proceed to page 555, procedure number UF- 7.0.0.
No Proceed to
page 483, SECTION 9- TMP PROBLEMS.
UF- 3.0.0 INADEQUATE UF PUMP OUTPUT
a) Per the Figure right, trace the UF Pump’s cable
to distribution board position “P22, UF-P”.
b) ENSURE the connector is placed
properly AND the cable is not damaged
see part c
c)
Place the machine in RINSE!
d) Ensure the external flow indicator’s ‘bob’ is moving up and down!
e) Per the Figure right, remove the tubing
from the UF Pump white INPUT (bottom)
nozzle. Is there flow through Filter #74
(Yes or No)?
Yes Flow through the filter!
See procedure number
UF- 4.0.0 (page 552).
No No flow! This may be the
problem! Replace UF Filter
#74.